Quality Control Description

Our measurements are automatically and manually examined by program staff daily. Trained on-site technicians service and maintain the instruments weekly, and are available to respond quickly for instrument problem resolution. These measures result in the collection of high quality data with very good capture rates.

When problems are discovered with the data, quality control abbreviations are set for each record. The quality control codes are separated by a semi-colon and appear both on the data graphs as well as in the data text file. A description of these quality control codes is listed below.

QC Code
Description
adUVA sensor problem, UVA data should not be used.
adcA/D converter failure, shadowband data and auxliary data should not be used.
atUVA dome temp is out of specifications, manufacture calibration is suspect.
baband is not adequately shadowing the diffuser, shadowband data should not be used.
bbsuspect bias applied to diffuse channel, total and diffuse shadowband data may be compromised.
bc1no des_factor in broadbands_calibrations for this serial_num, uvb broadband not calibrated.
bc2no erythemal yearly annual factor found for this serial_num, uvb broadband not calibrated.
bfbarometer failure, barometer data should not be used.
bgbroadband grounding problem. The data at night is not zero, and should be subtracted out. (formerly bd code)
bmband is not adequately shadowing the diffuse. The diffuse and direct data may be compromised
bsbroadband calibration data for this site not found, uvb broadband not calibrated.
btuvb broadband dome temp is out of specifications, calibrations is suspect
bxuvb broadband failure, broadband data should not be used.
bzbroadband dome is damaged or obscured, resulting in low transmission, broadband data should not be used.
c1Only beginning calibration file found, no interpolation done.
cddifferent shadowband calibration facility's files were used as the opening and closing calibration data.
cionly the initial/opening shadowband calibration file used in calibrating the uv-mfrsr instrument at this site.
cnshadowband calibration file not found for this site, data is not calibrated.
coshadowband calibration file for the uv-mfrsr instrument at this site is over 18 months old.
coyyankee angular correction file is being used on the uv-mfrsr instrument at this site. Only applies to shadowband data after year 2000.
cu[1..7]the shadowband calibration file had 0 in this channel's head or board gain. This shadowband channel was not calibrated.
cyyankee calibration file is being used on the the uv-mfrsr instrument at this site. Only applies to shadowband data after year 2000.
db[1..7]channel failure for listed channels caused by board damage, shadowband channel data should not be used.
dcno diffuse angular correction file found for this site, diffuse and total values not angularly corrected.
dddiffuser is damaged, shadowband channel voltages and calibrated irradiances may be compromised.
dh[1..7]channel failure for listed channels caused by loss of sensitivity of photodiode
drdiffuser is dirty, shadowband channel voltages and calibrated irradiances may be compromised.
ecaThe broadband sensor is erythemally corrected by using the constants provided by yankee. The constants are 0.141 if Solar zenith angle is <= 65 and 0.154 if > 65
edepply sensor problem, epply data should not be used.
efepply failure, epply data should not be used.
fppower failure
f[1..7]shadowband channel failure for listed channels, shadowband channel data should not be used for listed channels.
hcThe humidity is between 103 and 105 degrees.
hfhumidity failure, humidity data should not be used.
hh humidity is too high. Humidity data should not be used.
hthHead temperature is too high. Shadowband channel data may be compromised.
htlHead temperature is too low. Shadowband channel data may be compromised.
i[1..7]signal interference for listed channels. Shadowband channel data for listed channels may be compromised.
ldlicor problem, licor data should not be used.
lflicor sensor failure, licor data should not be used.
nbno bias was available to apply to diffuse channel data.
ngdiffuse and or direct values are negative, please see Data Processing Page for possible reasons.
nidNo head or board id found on instrument. Data cannot be corrected.
npno par correction data found for this site, par data is not corrected.
nsbA bias was created and used for the nasa instrument, by putting a dome over the instrument. The nighttime values were NOT used.
pdA par sensor problem, par sensor data should not be used.
pfA par sensor failure, par sensor data should not be used.
prA par sensor problem, the par sensor was obsured. The par data should not be used.
sl[1..n]problem with listed solar light instruments, solar light data should not be used.
snno angular correction file found for this site, data not angularly corrected.
socosine correction file for uv-mfrsr instrument is over a 18 months old.
sp[1..7]sensitivity problem for listed channels, lamp-calibrated irradiances should not be used, corresponding cosine-corrected voltages may be used for limited purposes.
s[1..7]signal saturated for listed channels. Shadowband channel data for listed channels may not be used.
tfair temperature failure, temperature data should not be used.
txBoard not synchronizing with the computer's time, timestamp and air mass may be off. If timing bias can be determined, the total horizontal and ancillary data may be usable.